How to understand GE Panametrics M-Series Probe part number

How to understand GE Panametrics M-Series Probe part number 




A
Model
M
Aluminum oxide moisture probe


B
Mount
1
Flange mount process connection with Viton O-ring, 300 series ss.
2
3/4 inch - 16 straight thread process connection with Viton O-ring, 300 series ss.
2N
3/4 inch - 16 Straight thread process connection mount meets NACE requirements.
7
auto insertion mechanism type 1 inch NPTM process connection 2800psig max press
Note- 1" fully opening ball valve required for isolation purposes
9
1-inch VCR gland mount; 316 SS
Note- Requires a 2891 sample cell


C
Hygrometer Designation
1
For use with MIOOO Hygrometer
2
For use with M2000 Hygrometer
3
For use with M3000 Hygrometer
4
For use with M4000 Hygrometer
L
Calibration range of -80 to +20 deg C dew/frost point with data to -110 deg C
Note- Option L for use with System 1/2, System 250, System 280,MISI, MIS2, MMS3, MMS35, MMS35IS, MTS5, PM880, MTS6


D
End cap
Y
Sintered stainless steel end cap/shield, 7 micron
R
Sintered stainless steel end cap/shield 100 micron
w
Sheet stainless steel end cap/shield
N
Nickel coated slip-on seal for 1” metal face seal process
Note- Probe shields are normally not used with metal face seal mounts (type M9), however, a shield will be used for shipping purposes to protect the sensor from damage. This shield should be removed at the time the probe is installed into the process.



E
Temperature Options
0
Temperature Options-None
T
Process Temperature Thermistor
Note- Temperature thermistor not available for the MMS35, MMS351S, MTS5 or MTS6


F
Calibration Options
0
Standard calibration, -80 TO +20C DEW/FROST POINT
Temp. with data to -110C DEW/FROST POINT
E
Extended calibration TO +60 DEG C DEW/FROST POINT
LE
Low end calibration -110 TO -50 DEG C DEW/FROST POINT
G
Standard calibration data extrapolated to +60 dew C dew point
Note- Low end calibration supersedes the standard calibration referenced above
s
Special calibration options     
z
Conditioning moisture probe for use in particular low moisture level applications

Note- If Calibration Option = E or LE, certification option G must equal O


G
Laboratory Services
0
Laboratory services none
1
Methanol aging of moisture probe
Note-  If extended range calibration option is chosen, then Methanol aging is not available
2
Cleaning of moisture probe for oxygen service


H
Certification
0
CERTIFICATION - NONE
1
NIST CERTIFICATON(National Institute of Standards and Technology)
2
NIST CERTIFICATION – Military STD 45662A
4
CERTIFICATE by NPL (National Physical Laboratory)


I
CRN Label
0
CRN labeling not required
1
CRN labeling, Include CPA-70 0-Ring
Z
Special


J
Options
0
None






FOR EXAMPLE if your part number is M2NLR-00-010-0



M-
Aluminum oxide moisture probe
2N
3/4 inch - 16 Straight thread process connection mount meets NACE requirements.
L
Calibration range of -80 to +20 DegC dew/frost point with data to -110 DegC
NOTE- Option L for use with System 1/2, System 250, System 280, MISI, MIS2, MMS3, MMS35, MMS35IS, MTS5, PM880, MTS6
R
Sintered stainless steel end cap/shield 100 micron
0
Temperature Options-None
0
Standard calibration, -80 TO +20C DEW/FROST POINT Temp. with data to -110C DEW/FROST POINT
0
Laboratory services none
1
NIST CERTIFICATON(National Institute of Standards and Technology)
0
CRN labeling not required
0
None


















Comments