AMETEK 933-How to do Auto Zero by using M933 Configurator software
![Image](https://blogger.googleusercontent.com/img/b/R29vZ2xl/AVvXsEgeWVWyrwerSW_FJ8lfhK4RTckU4Zp63SeKEEKWw8yyvHVOCaq8MMFof-HpIQCYBTREuDjsq-5lcIs99N1tsaXDMvxElwqydTRKOt1ZE3xLbYnidHgiMl_NG3aGBzBWcCoBBDBH9YY5i44/s1600/933.jpg)
STEP 1: After connecting the analyzer to M933 Configurator software Figure 1 STEP 2: Go to ‘edit’ ‘Device properties’ Figure 2 STEP 3: Go to ‘setup’ Figure 3 STEP 4: Go to ‘Gas Calibration’ Figure 4 NOTE-MAKE SURE THE FLOW CONTROL IN AUTO STEP 5: Press ‘AUTO ZERO’ and ‘OK’ After pressing ‘OK’ Press ‘Apply’ in M933 properties screen After pressing ‘Apply’ the below screen will appear Figure 5 NOTE: Press ‘YES’ to save the changes to EEPROM. STEP 6: Go to ‘STATUS’ ‘HOST STATE’ will change from ‘Operational to Operational – Performing Auto Zero’ Figure 6 NOTE: READING IN THE ANALYZER AND ALL OTHER PARAMETERS MAY DIFFER FROM RESPECTIVE TO EACH ANALYZER, IT IS JUST FOR REFERENCE STEP 7: G