AMETEK 933-How to do Auto Zero by using M933 Configurator software




STEP 1:
After connecting the analyzer to M933 Configurator software


Figure 1


  

STEP 2:
Go to ‘edit’
‘Device properties’

 Figure 2


STEP 3:
Go to ‘setup’


 Figure 3


STEP 4:
Go to ‘Gas Calibration’


 Figure 4
NOTE-MAKE SURE THE FLOW CONTROL IN AUTO


STEP 5:
Press ‘AUTO ZERO’ and ‘OK’
After pressing ‘OK’
Press ‘Apply’ in M933 properties screen
After pressing ‘Apply’ the below screen will appear



 Figure 5


NOTE: Press ‘YES’ to save the changes to EEPROM.

STEP 6:
Go to ‘STATUS’
‘HOST STATE’ will change from ‘Operational to Operational – Performing Auto Zero’


Figure 6
                                                                                                                                                            NOTE: READING IN THE ANALYZER AND ALL OTHER PARAMETERS MAY DIFFER FROM RESPECTIVE TO EACH ANALYZER, IT IS JUST FOR REFERENCE


STEP 7:
Go to ‘Analyzer Data’
In ‘Timing’ you can see Auto Zero calibration ending time
In Digital IO ‘solenoid’ will be ‘Zero’ 


Figure 7


STEP 8:
·         Calibration will end automatically after the ‘Calibration ends’ timing comes to -1 and your sample will be flow.
·         After calibration end
·         ‘Timing’ à ‘Cal Ends’ (sec) will be -1. As below
·         ‘Next column switch’ (sec) à will be showing next column switching time.
·         ‘Digital IO’ à solenoid will be showing current column

Figure 8

NOTE: NEXT COLUMN SWITCH TIMING, ANALYZER READING AND ALL OTHER PARAMETERS WILL BE DIFFER FROM RESPECTIVE TO EACH AND EVERY ANALYZER.

Disclaimer:
This procedure may not be same as the manual.
Some steps may be different
Image source from the internet
If any mistake or procedure left means kindly comment 



Comments