GE Panametrics Moisture Probe

GE Panametrics Moisture Probe

This moisture probe is also known as ALUMINIUM OXIDE MOISTURE SENSOR. An Aluminium probe is chemically treated to form a thin porous coating of aluminium oxide. The aluminium oxide is then coated with a thin, permeable layer of gold. . The gold layer and the aluminium probe form the anode and cathode. The gold layer is permeable to moisture and conductive. This forms the second electrode of a capacitor. The pores in the aluminum oxide layer absorb moisture from the gas stream in amounts directly proportional to the moisture content of the gas stream. The absorption of water molecules changes the capacitance of the sensor. The capacitance of the sensor is measured which is then converted to the moisture value. The radiuses of the pores in aluminium oxide layer make the sensor specific to water molecules.

















Types of Panametrics moisture probes
1. M Series     - Moisture and Temperature measurement
2. TF Series    - Moisture, Temperature and pressure measurement  
3.MISP Series - Moisture, Temperature and pressure measurement  

M Series      
M1 Probe Mount
o     Flange type probe mount with viton o-ring
         Used in applications like duckwork where you can’t tap a hole for 3/4” NPT or 3/4” straight thread
o    4 pins :
         2 for Al2O3 sensor
         2 for optional thermistor for temperature measurement

M2 Probe Mount
o   3/4”-16 straight thread with viton o-ring
o   4 pins :
         2 for Al2O3 sensor
         2 for optional thermistor for temperature measurement
M9 Probe Mount
o   1” VCR compatible VCR gland
o   Requires VCR metal gasket
o   4 pins :
         2 for Al2O3 sensor
         2 for optional thermistor for temperature measurement
o   Use W or R type shield.  Can remove shield for installation



TF Series  (Triple Function)   
TF-2 Probe Mount
o   3/4”-16 straight thread with viton o-ring
o   8 pins :
         2 for Al2O3 sensor
         4 pins for optional pressure transducer for pressure measurement
         2 for optional thermistor for temperature measurement
TF-9X Probe Mount
o   1” VCR compatible VCR gland
o   Requires VCR gasket
o   8 pins :
         2 for Al2O3 sensor
         2 for optional thermistor for temperature measurement
         4 pins for optional pressure transducer for pressure measurement

MISP Series  (Moisture Image Series Probe)
MISP Probe
o   Combination of TF-Probe and MISP module
         TF probe can be either TF-2 or TF-9 mount
o   Simple twisted pair probe cable
o   MISP module retains sensor calibration data
o   Incorporates 16bit A/D converter


Check this website for original post and more info :